Siemens - Maxum II Edition Gas Chromatograph - Safety Features
Read at HSEOpens the publisher’s page on hse.gov.uk
Summary
HSE alerts gas chromatograph users to an explosion that caused life-changing injuries during return to service of a Siemens Maxum II. Hydrogen had accumulated while the unit was on hold, and automatic purging was not re-engaged. Users should review purge controls and discuss operating arrangements with manufacturers.